• Source 0 to 24mA with 0.01mA resolution • Measure 0 to 50mA (0.1mA resolution) • Select Type K, J, or T thermocouple for precision calibrations in °F/°C or mV • Output 0 to 1999mV or 0 to 10.00V • Measure 0 to 19.99V • Single/continuous step function for Voltage and Current • Self-contained 24V loop power supply drives current loads up to 1000Ohm • 5 stored calibration values of each type for fast calibrations • Complete with six AA batteries and calibration cable with spade lug terminals
Specs:
Specifications
Range
Accuracy (of reading)
DC Source
Current
0 to 24mA, -25% to +125%
±(0.075% + 1 digit) or ± 3 digits, whichever is greater
Voltage
0 to 10V
±(0.075% + 1 digit) or ± 3 digits, whichever is greater
Type J
-58 to 1830°F (-50 to 1000°C)
±(0.15% + 1.8°F or 1°C)
Type K
-58 to 2498°F (-50 to 1370°C)
±(0.15% + 1.8°F or 1°C)
Type T
-184 to 752°F (-120 to 400°C)
±(0.15% + 1.8°F or 1°C)
Loop Power
24V
±(10μV ± 1 digit)
Max. Load
1000
@ 24mA
DC Measure
Current
0 to 50mA, -25% to +230%
±(0.075% + 1 digit) or ± 3 digits, whichever is greater
Voltage
0 to 20V
±(0.075% + 1 digit) or ± 3 digits, whichever is greater
Type J
-58 to 1830°F (-50 to 1000°C)
±(0.15% + 1.8°F or 1°C)
Type K
-58 to 2498°F (-50 to 1370°C)
±(0.15% + 1.8°F or 1°C)
Type T
-184 to 752°F (-120 to 400°C)
±(0.15% + 1.8°F or 1°C)
Dimensions/Weight
3.8 x 4.7 x 2.0" (96 x 118 x 45mm)/12oz (340g)