Features:
These easy-to-use, microprocessor-based meters feature hold, ATC, recorder output, and electrode diagnostics. The convenient pull-out quick reference card keeps procedures handy at all times. The large, dual-display LCD provides excellent visibility—even from a distance. Meters accept electrodes with a BNC connector.
Measures pH, temperature, and relative mV
Five-point pH calibration with automatic buffer recognition for both NIST (pH 1.68,4.01, 6.86, 9.18, 12.45) and USA (pH 1.68, 4.01, 7.00, 10.01,12.45) pH buffers
Automatic temperature compensation maintains reading accuracy even in fluctuating temperatures.
Includes all in one temperature /single junction pH electrode (#35811-71) and AC adapter.
Specifications
| Range |
pH |
0.00 to 14.00 pH |
| mV, relative |
±1000 mV |
| Temperature |
0 to 100°C |
| Resolution |
pH |
0.01 pH |
| mV, relative |
0.1 mV within ±199.9 mV; 1 mV beyond ±199.9 mV |
| Temperature |
0.1°C |
| Accuracy |
pH |
±0.01 pH |
| mV, relative |
0.2 mV within ±199.9 mV; 2 mV beyond ±199.9 mV |
| Temperature |
±0.3°C |
| Buffer recognition |
5 points (select NIST or USA buffer sets) |
| Temp compensation |
Manual or automatic from 0 to 100.0°C |
| Recorder output |
± 1999 mV, 1 mV/mV unit |
| Input impedance |
>1012W |
| Slope control |
80 to 120% |
| Display |
Dual LCD with mode annunciators |
| Power |
110 VAC |
| Operating temperature |
0 to 50°C |
| Dimensions |
9"L x 7"W x 2 3/8"H |
|